Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: Results from the Micra atrioventricular feasibility studies

Larry A. Chinitz(NYU Langone Health), Philippe Ritter(Hôpital Cardiologique du Haut-Lévêque), Surinder Kaur Khelae(National Heart Institute), Saverio Iacopino, Christophe Garweg(Universitair Ziekenhuis Leuven), Maria Grazia-Bongiorni(Azienda Ospedaliera Universitaria Pisana), Petr Neužil(Na Homolce Hospital), Jens Brock Johansen(Odense University Hospital), Lluı́s Mont(Universitat de Barcelona), Efraín González(Baptist Hospital of Miami), Venkata Sagi, Gábor Zoltán Duray, Nicolas Clémenty(Centre Hospitalier Universitaire de Tours), Todd Sheldon(Medtronic (United States)), Vincent Splett(Medtronic (United States)), Kurt Stromberg(Medtronic (United States)), Nicole Wood(Medtronic (United States)), Clemens Steinwender(Paracelsus Medical University)
Heart Rhythm
May 11, 2018
Cited by 162Open Access
Full Text

Abstract

BackgroundMicra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling atrioventricular (AV) synchronous pacing.ObjectiveThe purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing.MethodsThe Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms.ResultsA total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0–41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%–90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P <.001) during AV algorithm pacing compared to VVI in high-degree block patients, whereas AVS was maintained in patients with intrinsic conduction.ConclusionAccelerometer-based atrial sensing is feasible and significantly improves AVS in patients with AV block and a single-chamber leadless pacemaker implanted in the right ventricle. Micra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling atrioventricular (AV) synchronous pacing. The purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing. The Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms. A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0–41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%–90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P <.001) during AV algorithm pacing compared to VVI in high-degree block patients, whereas AVS was maintained in patients with intrinsic conduction. Accelerometer-based atrial sensing is feasible and significantly improves AVS in patients with AV block and a single-chamber leadless pacemaker implanted in the right ventricle.


Related Papers

No related papers found

Powered by citation graph analysis