A label-free immunosensor array using single-chain antibody fragments

Natalija Backmann(University of Basel), Christian Zahnd(University of Basel), François Huber(University of Basel), Alexander Bietsch(University of Basel), Andreas Plückthun(University of Basel), H.P. Lang(University of Basel), H.‐J. Güntherodt(University of Basel), Martin Hegner(University of Basel), Christoph Gerber(University of Basel)
Proceedings of the National Academy of Sciences
September 28, 2005
Cited by 277

Abstract

We report a microcantilever-based immunosensor operated in static deflection mode with a performance comparable with surface plasmon resonance, using single-chain Fv (scFv) antibody fragments as receptor molecules. As a model system scFv fragments with specificity to two different antigens were applied. We introduced a cysteine residue at the C terminus of each scFv construct to allow covalent attachment to gold-coated sensor interfaces in directed orientation. Application of an array enabled simultaneous deflection measurements of sensing and reference cantilevers. The differential deflection signal revealed specific antigen binding and was proportional to the antigen concentration in solution. Using small, oriented scFv fragments as receptor molecules we increased the sensitivity of microcantilevers to approximately 1 nM.