A Review of Methods for Correction of Intensity Inhomogeneity in MRI

Uro Vovk(University of Ljubljana), Franjo Pernuš(University of Ljubljana), B. Likar(University of Ljubljana)
IEEE Transactions on Medical Imaging
March 1, 2007
Cited by 837

Abstract

Medical image acquisition devices provide a vast amount of anatomical and functional information, which facilitate and improve diagnosis and patient treatment, especially when supported by modern quantitative image analysis methods. However, modality specific image artifacts, such as the phenomena of intensity inhomogeneity in magnetic resonance images (MRI), are still prominent and can adversely affect quantitative image analysis. In this paper, numerous methods that have been developed to reduce or eliminate intensity inhomogeneities in MRI are reviewed. First, the methods are classified according to the inhomogeneity correction strategy. Next, different qualitative and quantitative evaluation approaches are reviewed. Third, 60 relevant publications are categorized according to several features and analyzed so as to reveal major trends, popularity, evaluation strategies and applications. Finally, key evaluation issues and future development of the inhomogeneity correction field, supported by the results of the analysis, are discussed.