Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II System

Erland J. Erlandsen(Regionshospitalet Viborg), Else Randers(Regionshospitalet Viborg), Jens Kristensen(Regionshospitalet Viborg)
Scandinavian Journal of Clinical and Laboratory Investigation
January 1, 1999
Cited by 259

Abstract

The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometric immunoassay for measuring serum cystatin C, was evaluated on the Dade Behring Nephelometer II. The assay time was 6 min and the throughput was 75 samples per hour. The sample volume was 40 microL and the measuring range was 0.25-7.90 mg/L. Imprecision studies revealed within-run CVs < 1.8% and between-run CVs < 1.8% in the concentration range 0.87-4.63 mg/L. Recovery was 92.4-101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemoglobin < 1.0 mmol/ L, bilirubin <512 micromol/L and Intralipid <20 g/L. Stability of cystatin C in serum was 7 days at temperatures from 20 degrees C to 20 degrees C and 6 months at -80 degrees C. Measurements of cystatin C in heparin-plasma and EDTA-plasma did not differ significantly from cystatin C measured in serum. Fifty patient samples run on the Dade Behring Nephelometer II (y) were compared to the Dako Cystatin C assay (x). The Passing-Bablok regression analysis revealed y = 1.105x - 0.340. In conclusion, the Dade Behring N Latex Cystatin C assay was precise and correlated with the Dako Cystatin C assay.


Related Papers

No related papers found

Powered by citation graph analysis