The Nellcor N‐101 pulse oximeter
J. A. TYTLER(St George's Hospital), H. F. SEELEY(St George's Hospital)
Cited by 32Open Access
Abstract
The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.