The Nellcor N‐101 pulse oximeter

J. A. TYTLER(St George's Hospital), H. F. SEELEY(St George's Hospital)
Anaesthesia
March 1, 1986
Cited by 32Open Access
Full Text

Abstract

The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.