Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors

Boris Kovatchev(University of Virginia Health System), Linda Gonder‐Frederick(University of Virginia Health System), Daniel J. Cox(University of Virginia Health System), William L. Clarke(University of Virginia Health System)
Diabetes Care
August 1, 2004
Cited by 311Open Access
Full Text

Abstract

OBJECTIVE: The objective of this study was to introduce continuous glucose-error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator. RESEARCH DESIGN AND METHODS: We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point-error grid analysis (P-EGA) evaluates the sensor's accuracy in terms of correct presentation of BG values and 2) rate-error grid analysis (R-EGA) assesses the sensor's ability to capture the direction and rate of BG fluctuations. RESULTS: CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels. CONCLUSIONS: Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.


Related Papers

No related papers found

Powered by citation graph analysis