Elimination efficiency of different reagents for the memory effect of mercury using ICP-MS

Yufeng Li(Institute of High Energy Physics), Chunying Chen(Chinese Academy of Sciences), Yuanyuan Li(Chinese Academy of Sciences), Jin Sun(Institute of High Energy Physics), Jiangxue Wang(Chinese Academy of Sciences), Yuxi Gao(Institute of High Energy Physics), Yuliang Zhao(Institute of High Energy Physics), Zhifang Chai(Chinese Academy of Sciences)
Journal of Analytical Atomic Spectrometry
November 16, 2005
Cited by 351

Abstract

The present study was carried out to solve the problems of long washout time and non-linear calibration curves encountered in mercury analysis using inductively coupled plasma mass spectrometry. Comparisons of the washing efficiency for different reagents to eliminate the mercury memory effect were made. Of all the selected washing reagents, mercapto reagents such as 2-mercaptoethanol and L-cysteine could efficiently clean the instrument. Taking the toxicity and odor of mercaptoalkanol into account, L-cysteine was preferred as the most suitable washing reagent and was added into the standard and sample solutions. A good linear calibration curve was obtained with the correlation coefficient of 0.9999. The detetion limit for addition of L-cysteine was 0.024 μg l−1. Addition of 0.18% L-cysteine into the sample solution also facilitates the washout of mercury even using deionized water. The recoveries of mercury in certified reference materials, i.e. human hair and dogfish muscle, were 97.2% and 98.3%, respectively, when using 0.18% L-cysteine in the sample solutions.