Integrated parallel reception, excitation, and shimming (iPRES)

Hui Han(Duke University), Allen W. Song(Duke University), Trong‐Kha Truong(Duke University)
Magnetic Resonance in Medicine
April 29, 2013
Cited by 79Open Access
Full Text

Abstract

PURPOSE: To develop a new concept for a hardware platform that enables integrated parallel reception, excitation, and shimming. THEORY: This concept uses a single coil array rather than separate arrays for parallel excitation/reception and B0 shimming. It relies on a novel design that allows a radiofrequency current (for excitation/reception) and a direct current (for B0 shimming) to coexist independently in the same coil. METHODS: Proof-of-concept B0 shimming experiments were performed with a two-coil array in a phantom, whereas B0 shimming simulations were performed with a 48-coil array in the human brain. RESULTS: Our experiments show that individually optimized direct currents applied in each coil can reduce the B0 root-mean-square error by 62-81% and minimize distortions in echo-planar images. The simulations show that dynamic shimming with the 48-coil integrated parallel reception, excitation, and shimming array can reduce the B0 root-mean-square error in the prefrontal and temporal regions by 66-79% as compared with static second-order spherical harmonic shimming and by 12-23% as compared with dynamic shimming with a 48-coil conventional shim array. CONCLUSION: Our results demonstrate the feasibility of the integrated parallel reception, excitation, and shimming concept to perform parallel excitation/reception and B0 shimming with a unified coil system as well as its promise for in vivo applications.