Pulsed CO<sub>2</sub> laser tissue ablation: Effect of tissue type and pulse duration on thermal damage

Joseph T. Walsh(Massachusetts General Hospital), Thomas J. Flotte(Massachusetts General Hospital), R. Rox Anderson(Harvard University Press), Thomas F. Deutsch(Harvard University Press)
Lasers in Surgery and Medicine
January 1, 1988
Cited by 375

Abstract

Tissue removal by infrared lasers is accompanied by thermal damage to nonablated tissue. The extent of thermal damage can be controlled by a choice of laser wavelength, irradiance, and exposure duration. The effect of exposure duration has been studied in vivo by using CO2 lasers with pulse widths that vary from 2 microseconds to 50 msec. Pulse widths of 50 msec, typical of a shuttered, continuous-wave CO2 laser, produce damage regions 750 micron wide in normal guinea pig skin; the use of a 2-microseconds-long pulse reduced this damage zone to as little as 50 micron. Using 2-microseconds-long pulses, in vitro studies showed that the minimum zone of thermal damage varied significantly with tissue type. The thermal denaturation of these tissues has been studied and correlated with damage. The effect of denaturation temperature and pulse duration on the width of the damage zone is explained by a simple model.


Related Papers

No related papers found