Solvent Microextraction into a Single Drop

Michael A. Jeannot(University of Alberta), Frederick F. Cantwell(University of Alberta)
Analytical Chemistry
July 1, 1996
Cited by 1,273

Abstract

An analytical technique is described which combines solvent extraction with gas chromatographic (GC) analysis in a simple and inexpensive apparatus involving very little solvent consumption. A small drop (8 μL) of a water-immiscible organic solvent, containing an internal standard, is located at the end of a Teflon rod which is immersed in a stirred aqueous sample solution. After the solution has been stirred for a prescribed period of time, the probe is withdrawn from the aqueous solution, and the organic phase is sampled with a microsyringe and injected into the GC for quantification. The observed rate of solvent extraction is in good agreement with a convective-diffusive kinetic model. Analytically, the relative standard deviation of the method is 1.7% for a 5.00-min extraction of the analyte 4-methylacetophenone into n-octane.


Related Papers

No related papers found