A Prospective Diagnostic Accuracy Study of 18F-Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography, Multidetector Row Computed Tomography, and Magnetic Resonance Imaging in Primary Diagnosis and Staging of Pancreatic Cancer

Saila Kauhanen(University of Turku), Gaber Komar(University of Turku), Marko Seppänen(University of Turku), Kirsti Dean(Varsinais-Suomen Sairaanhoitopiiri), Heikki Minn(University of Turku), Sami Kajander(University of Turku), Irina Rinta‐Kiikka(Tampere University Hospital), Kalle Alanen(Turku University Hospital), Ronald Borra(University of Turku), Pauli Puolakkainen(University of Turku), Pirjo Nuutila(University of Turku), Jari Ovaska(University of Turku)
Annals of Surgery
November 23, 2009
Cited by 325

Abstract

In Brief Objective: To prospectively compare the accuracy of combined positron emission tomography/computed tomography using 18F-fluorodeoxyglucose (FDG-PET/CT), multidetector row computed tomography (MDCT), and magnetic resonance imaging (MRI) in the evaluation of patients with suspected pancreatic malignancy. Summary Background Data: FDG-PET/CT imaging is increasingly used for staging of pancreatic cancer. Preliminary data suggest a significant influence of FDG-PET/CT on treatment planning, although its role is still evolving. Methods: Thirty-eight consecutive patients with suspicion of pancreatic malignancy were enrolled. Patients underwent a protocol including FDG-PET/CT, MDCT, and MRI combined with magnetic resonance cholangiopancreatography, all of which were blindly evaluated. The findings were confirmed macroscopically at operation and/or by histopathologic analysis (n = 29) or follow-up (n = 9). Results of TNM classification of different imaging methods were compared with clinical TNM classification. Results: Pancreatic adenocarcinoma was diagnosed in 17 patients, neuroendocrine tumor in 3, mass-forming pancreatitis in 4, cystic lesion in 6, and fibrosis in 2. Six patients had a finding of a normal pancreas. The diagnostic accuracy of FDG-PET/CT for pancreatic malignancy was 89%, compared with 76% and 79% for MDCT and MRI, respectively. In the differential diagnosis of suspected malignant biliary stricture at endoscopic retrograde cholangiopancreaticography (n = 21), FDG-PET/CT had a positive predictive value of 92%. In 17 patients with advanced pancreatic adenocarcinoma, FDG-PET/CT had a sensitivity of 30% for N- and 88% for M-staging. Both MDCT and MRI had sensitivities of 30% for N- and 38% for M-staging. Furthermore, the clinical management of 10 patients (26%) was altered after FDG-PET/CT. Conclusion: FDG-PET/CT was more sensitive than conventional imaging in the diagnosis of both primary pancreatic adenocarcinoma and associated distant metastases. In contrast, the sensitivity of FDG-PET/CT was poor in detecting local lymph node metastasis, which would have been important for an assessment of resectability. We recommend the use of FDG-PET/CT in the evaluation of diagnostically challenging cases, especially in patients with biliary strictures without evidence of malignancy in conventional imaging. The aim of the study was to prospectively compare 18F-fluorodeoxyglucose positron emission tomography or computed tomography, multidetector row computed tomography, and magnetic resonance imaging in the diagnosis of pancreatic cancer. In group of 38 patients, 18F-fluorodeoxyglucose positron emission tomography or computed tomography was more sensitive than conventional imaging in the diagnosis of both primary pancreatic adenocarcinoma and associated distant metastases.


Related Papers

No related papers found

Powered by citation graph analysis