A method for incorporating organ motion due to breathing into 3D dose calculations

Anthony E. Lujan(University of Michigan), Edward W. Larsen(University of Michigan), James M. Balter(University of Michigan), Randall K. Ten Haken(University of Michigan)
Medical Physics
May 1, 1999
Cited by 506Open Access
Full Text

Abstract

A method is proposed that incorporates the effects of intratreatment organ motion due to breathing on the dose calculations for the treatment of liver disease. Our method is based on the convolution of a static dose distribution with a probability distribution function (PDF) which describes the nature of the motion. The organ motion due to breathing is assumed here to be one-dimensional (in the superior-inferior direction), and is modeled using a periodic but asymmetric function (more time spent at exhale versus inhale). The dose distribution calculated using convolution-based methods is compared to the static dose distribution using dose difference displays and the effective volume (Veff) of the uninvolved liver, as per a liver dose escalation protocol in use at our institution. The convolution-based calculation is also compared to direct simulations that model individual fractions of a treatment. Analysis shows that incorporation of the organ motion could lead to changes in the dose prescribed for a treatment based on the Veff of the uninvolved liver. Comparison of convolution-based calculations and direct simulation of various worst-case scenarios indicates that a single convolution-based calculation is sufficient to predict the dose distribution for the example treatment plan given.


Related Papers

No related papers found

Powered by citation graph analysis