An AutoAnalyzer Procedure for Serum Iron and Total Iron-Binding Capacity, with Use of Ferrozine
Hugh Y. Yee(Hutzel Women's Hospital), Anastasia Zin(Hutzel Women's Hospital)
Cited by 35
Abstract
Abstract An AutoAnalyzer method has been developed in which the recently introduced reagent for iron, 3-(2-pyridyl)-5,6-bis(4-phenylsulfonic acid)-1,2,4-triazine ("ferrozine"), is used. With the Technicon colorimeter or Gilford 300-N spectrophotometer in use, sampling rates may be 30/h or 50/h, respectively. Sample volume is 0.8 ml, and measurements are made at 560 nm. A procedure for measuring total iron-binding capacity is also described.
Related Papers
No related papers found